Endotension as a result of pressure transmission through the graft following endovascular aneurysm repair--an in vitro study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 14532877.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
endovascular aneurysm repair (EVAR) significantly reduces, but does not abolish aneurysm sac pressure, possibly because of trans-fabric transmission. to investigate how blood pressure is transmitted through different types of grafts into the aneurysm sac. experimental study, in vitro. a latex aneurysm was inserted into an in vitro circulation model. The systemic mean pressure (SPmean) was varied from 50 to 120 mmHg. The grafts used for aneurysm exclusion were: thin wall polyethylene (PE), thick wall polyethylene (PE) and thin wall ePTFE. Mean aneurysm sac pressure (ASPmean) was measured, as was pulse pressure (ASPpulse). at an SPmean of 70 mmHg, the ASPmean was 34 +/- 0.8 mmHg (polyethylene knitted, thick wall), 30 +/- 1.0 mmHg (polyethylene woven, thin wall), and 17 +/- 0.6 mmHg (thin wall ePTFE). The ASPmean increased with SPmean, the relationship depending on the graft material. Stiffer grafts were associated with lower ASPmean and ASPpulse (p<0.001). the relationship between aneurysm sac mean pressure and systemic pressure (SP) depends on the graft material. These data highlights the need for further studies regarding endotension.