Haemodynamic properties of PTFE femoropopliteal bypass grafts as determined by a new magnetic resonance technique.

Sørensen, H D; Paaske, W P; Pedersen, E M · Eur J Vasc Endovasc Surg · 2003

biomechanical · Level V

Where this comes from

Abstract

To apply the automatic three dimensional paraboloid (3DP) method for measurement of wall shear stress (WSS), blood flow, blood velocity, and cross sectional graft lumen area to magnetic resonance phase contrast velocity mapping data acquired from polytetrafluoroethylene (PTFE) suprageniculate 6 mm femoropopliteal bypass grafts to determine the biomechanical and haemodynamic properties immediately and three years after implantation. Two patient cohorts, each of ten patients, with intermittent claudication were investigated by the 3DP method, one group (A) within 36 h after implantation of a femoropopliteal graft and the other group (B) 35 months after implantation. The mean blood flow was 5.9 ml/s (SD: 1.6), and WSS at peak systole was 2.2 N/m2 (SD: 0.6) 5 cm upstream to the distal anastomosis for group A, and 3.7 ml/s (SD: 1.9) and 1.9 N/m2 (SD: 0.4) for Group B. The WSS varied according to the angular position of the graft circumference. The implanted grafts were not circular but had a slightly elliptical circumference with a greater anterior/posterior diameter. The 3DP method is a precise tool with subpixel resolution for determining the biomechanical and haemodynamic properties of implanted PTFE grafts, and it can be used to assess graft function immediately after implantation. It is potentially applicable for routine graft surveillance.