Failed meniscus repair.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 14551564.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This report describes problems arising after an arthroscopic meniscal refixation using the biodegradable Clearfix screw system (Innovasive Devices, Marlborough, MA). A serious problem arose in the form of migration of a meniscal screw that had been inserted in accordance with the manufacturer's instructions. Dislocation of the screw led to pain on the medial aspect of the knee joint and irritation of the infrapatellar branch of the saphenous nerve. The case reported here is a first analysis of the potential complications associated with the use of this meniscal refixation system.
Medical subject headings
- Absorbable Implants
- Bone Screws
- Foreign-Body Migration
- Menisci, Tibial
- Postoperative Complications
Anatomy
- tibia