Fabric tears as a new cause of type III endoleak with Ancure endograft.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 14560241.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We present two case reports of type IIIb endoleak. One was due to fabric erosion caused by placement of a stent (Wallstent; W. L. Gore & Associates, Flagstaff, Ariz) after endovascular aneurysm repair; the other arose spontaneously. In both cases, an Ancure endograft (Guidant/EVT, Menlo Park, Calif) was placed. In case 1, a large endoleak developed 36 months after uncomplicated endovascular treatment of an abdominal aortic aneurysm. In case 2, endoleak developed 30 months after a complicated procedure. In both cases, two Wallstents were used to treat type I endoleak and limb kinking in the first postoperative months. One type III endoleak was within the endograft at the level of the stents. To our knowledge, these are the first type III endoleaks reported in association with Ancure endografts. Placement of Wallstents inside endografts is of concern, and another indication for close monitoring during follow-up.
Medical subject headings
- Aortic Aneurysm, Abdominal
- Blood Vessel Prosthesis Implantation
- Prosthesis Failure
- Stents