Transient dystonia following magnetic resonance imaging in a patient with deep brain stimulation electrodes for the treatment of Parkinson disease. Case report.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 14567615.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Data from previous studies have shown that magnetic resonance (MR) imaging of the head can be performed safely in patients with deep brain stimulators. The authors report on a 73-year-old patient with bilaterally implanted deep brain electrodes for the treatment of Parkinson disease, who exhibited dystonic and partially ballistic movements of the left leg immediately after an MR imaging session. Such dystonic or ballistic movements had not been previously observed in this patient. In the following months, this focal movement disorder resolved completely. This case demonstrates the possible risks of MR imaging in patients with deep brain stimulators.
Medical subject headings
- Brain
- Dystonia
- Electric Stimulation Therapy
- Magnetic Resonance Imaging
- Parkinson Disease