Comparing a mechanical device with medicinal leeches for treating venous congestion.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 14595279.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare mechanical device therapy to medicinal leech therapy for treating venous congestion. Fasciocutaneous flaps of 9 x 7-cm (porcine model, n = 11) were elevated and the associated veins clamped for 15 hours. Device-treated flaps (n = 6) were treated with up to 3 devices/flap. Leech-treated flaps (n = 5) were treated with 75 leeches/flap (5 leeches/hour). Treatment evaluation parameters included: blood volume removed, skin color, surface perfusion, tissue oxygen tension, and endpoint histology. No significant differences were seen in blood volume removed, surface perfusion, and oxygen tension. However, significant differences were seen in skin color change over time. Histological assessment showed mild congestion in 5/5 leech-treated flaps and 2/6 device-treated flaps. Device therapy was comparable to leech therapy for decongesting a 9 x 7-cm flap over a 15-hour period. However, digital images and significant differences in skin color change over time demonstrate that device therapy promotes more global decongestion versus leech therapy.
Medical subject headings
- Laser-Doppler Flowmetry
- Leeching
- Spectrophotometry
- Surgical Flaps
- Venous Thrombosis