Bonding strength of the apatite layer formed on glass-ceramic apatite-wollastonite-polyethylene composites.

Juhasz, J A; Best, S M; Kawashita, M; Miyata, N; Kokubo, T; Nakamura, T; Bonfield, W · J Biomed Mater Res A · 2003

basic_science · Level V

Where this comes from

Abstract

Bioactive glass-ceramic apatite-wollastonite (A-W) has been incorporated into polyethylene in particulate form to create new bioactive composites for potential maxillofacial applications. The effects of varying the volume fraction of glass-ceramic A-W filler and the glass-ceramic A-W particle size were investigated by measuring the bonding strength of the bonelike apatite layer formed on the surface of glass-ceramic A-W-polyethylene composites. The bonding strength was evaluated via a modified ASTM C-333 standard in which a tensile stress was applied to the substrate and the strength of the bioactive layer was compared with that formed on commercially available hydroxyapatite-polyethylene composite samples, HAPEX. The composites demonstrated greater bonding strength with increased filler content and reduced filler particle size (maximum 6.9 +/- 0.5 MPa) and a marginally greater bonding strength as compared with HAPEX (2.8 +/- 0.5 MPa), when glass-ceramic A-W-polyethylene composite samples with the same filler content were tested. The higher bonding strength of the apatite layer formed on the A-W-polyethylene composite samples suggests that, in addition to maxillofacial applications, these composites might also be utilized in applications involving higher levels of load bearing.