Intermittent-mode CT fluoroscopy-guided biopsy of the lung or upper abdomen with breath-hold monitoring and feedback: system development and feasibility.

Carlson, Stephanie K; Felmlee, Joel P; Bender, Claire E; Ehman, Richard L; Classic, Kelly L; Hu, Houchun H; Hoskin, Tanya L · Radiology · 2003

case_series · Level IV

Where this comes from

Abstract

A bellows-based breath-hold monitoring and feedback system was developed and evaluated for use in intermittent-mode computed tomographic (CT) fluoroscopy-guided biopsy procedures in the lung or upper abdomen. The bellows system is described, and its feasibility is demonstrated in studies with a respiratory phantom and human volunteers. Results are reported for seven patients who underwent bellows-assisted biopsy. Breath-hold monitoring and feedback with the bellows system allow the patient to perform reliable breath holding at a preselected level. This optimizes intermittent-mode CT fluoroscopy-guided biopsies by allowing consistent visualization of the target lesion throughout the procedure.

Medical subject headings