Recurrent dislocation of total knee arthroplasty: an unusual cause.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 14658114.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report an unusual case of recurrent dislocation of posterior stabilized total knee arthroplasty (TKA) secondary to fracture of the polyethylene insert. Recurrent dislocation occurred 21 months after primary TKA. Fractured polyethylene insert was noted as an abnormal radiolucent shadow on postoperative radiographs and was confirmed by an arthrogram. At revision surgery, disintegration of the insert, including the tibial post, was found. We recommend careful scrutiny of the integrity of the polyethylene insert at the time of follow-up evaluation in cases of TKA dislocation.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Postoperative Complications
Anatomy
- knee