Clinical evaluation of a breathing protocol for PET/CT.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 14676976.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to assess the frequency and severity of respiration-induced curvilinear respiration artifacts (RICA) on co-registered positron emission tomography/computed tomography (PET/CT) images acquired on a combined PET/CT scanner before and after modifying the respiration protocol for CT scanning, with retrospective analysis of two groups of 100 patients each, before and after implementing a respiration protocol with breath-hold (BH) in the normal expiration position for the acquisition of the CT images. The CT data were used as attenuation map and for image co-registration. A ranking of co-registered PET/CT and PET images (including maximum intensity projection) was done by two observers in consensus using a scale from 0 to 3. Zero indicated that no RICA was visible and 1, 2, and 3 described artifact with increasing severity. A significant difference in RICA occurrence was found between the two groups ( p<0.0001). There was a 45% decrease of artifact frequency when using the normal expiration protocol and a 68% decrease of grade-2 and grade-3 artifacts ( p=0.004). The results of this study suggest that BH during the normal expiration position for CT scanning can be recommended to reduce the occurrence and the severity of RICA on PET/CT.