A novel expiratory circuit for recovery of perfluorocarbon liquid during partial liquid ventilation.

Dunster, Kimble Robert; Davies, Mark William · Intensive Care Med · 2004

basic_science · Level V

Where this comes from

Abstract

To determine whether perfluorocarbon liquid can be condensed from gases containing perfluorocarbon vapour and whether the amount recovered varies with background flow rate. Bench-top experimental study in a neonatal laboratory. The expiratory limb of a standard ventilator circuit set-up was mimicked, with the addition of a chilled water jacket (Liebig) condenser. Perfluorocarbon vapour was passed through the circuit at a number of flow rates. Perfluorocarbon vapour was passed through the circuit and the percentage recovery of liquid measured. More than 60% of the perfluorocarbon vapour was recovered at all flow rates (1, 2, 5 and 10 l/min), with significantly higher recovery obtained (up to 74%) at low flow rates (1 l/min). Using a simple condenser, more than 60% of perfluorocarbon liquid can be recovered without altering the function of an expiratory limb of a ventilator circuit.