e-PTFE ventricular shunt catheters.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1470327.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The perfect polymer surface for implantation should resist the adhesion of bacteria while producing no reaction in surrounding tissues. Silicone elastomer is a common material used for medical implantation devices. This material is superior to most available for implantation because of its flexibility and low, but detectable, tissue reaction. We have evaluated a flexible, expanded polytetrafluoroethylene (e-PTFE) ventricular catheter for cerebrospinal fluid shunting and have found that although e-PTFE is safe for cerebral implantation, the porosity of the material allows tissue ingrowth that obstructs the lumen. Our limited clinical evaluation of 5-microns internodal distance e-PTFE revealed that this open-cell structure, allowing tissue ingrowth, apparently is a poor candidate to replace silicone elastomer in cerebrospinal fluid shunting devices. However, further investigation may provide a polymer superior to silicone elastomer to create a cerebrospinal fluid shunt.
Medical subject headings
- Polytetrafluoroethylene
- Silicone Elastomers
- Ventriculoperitoneal Shunt