Abstract
The strength-duration curve is a plot of the threshold current (I) versus pulse duration (d) required to stimulate excitable tissue. On this curve are two points: 1) rheobase (b) and 2) chronaxie (c). Rheobase is the threshold current for an infinitely long-duration stimulus. Chronaxie, the excitability constant, is the duration of a pulse of current of twice rheobasic strength. The mathematical expression for the strength-duration curve is I = b(1 + c/d). Although there are many published values for chronaxie for various excitable tissues, the range of variability for a given tissue type is quite large. This paper identifies five factors that can affect the accuracy of chronaxie measurement and shows that the most reliable values can be obtained with a rectangular pulse delivered from a constant-current source.
Medical subject headings
- Action Potentials
- Cell Membrane
- Chronaxy
- Electric Stimulation
- Energy Transfer
- Membrane Potentials