Time-frequency detection and analysis of wheezes during forced exhalation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 14723508.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The objective of the present work was to detect and analyze wheezes by means of a highly sensitive time-frequency algorithm. Automatic measurements were compared with clinical auscultation for forced exhalation segments from 1.2 to 0 liters/second (l/s). Sensitivities between 100% and 71%, as a function of flow level related to wheezing segments detection, were achieved. Time-frequency wheeze parameters were measured for the flow range from 1.2 to 0.2 l/s. Wheezes were detected in both analyzed groups; asthmatics (N = 16) and control subjects (N = 15). Significant differences between groups were found for the mean number of wheezes detected at basal condition (p = 0.0003). Frequency parameter differences were also significant (0.0112 < p < 0.0307). All these parameters were also studied after applying a bronchodilator drug (Terbutaline). Significant differences between patient groups were found when studying the changes in the number of wheezes for each patient (p = 0.0195). Finally, limited bandwidth parameters, which measure the bronchodilator response, were also studied.
Medical subject headings
- Algorithms
- Asthma
- Auscultation
- Diagnosis, Computer-Assisted
- Pattern Recognition, Automated
- Respiratory Sounds
- Sound Spectrography