A randomized controlled trial of the Arctic Sun Temperature Management System versus conventional methods for preventing hypothermia during off-pump cardiac surgery.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 14742358.
- Also identified by DOI 10.1213/01.ANE.0000096242.06561.C0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this trial we compared the hypothermia avoidance abilities of the Arctic Sun Temperature Management System (a servo-regulated system that circulates temperature-controlled water through unique energy transfer pads adherent to the patient's body) with conventional temperature control methods. Patients undergoing off-pump coronary artery bypass (OPCAB) surgery were randomized to either the Arctic Sun System alone (AS group) or conventional methods (control group; increased room temperature, heated IV fluids, convective forced air warming system) for the prevention of hypothermia (defined by a temperature <36 degrees C). The AS group had nasopharyngeal temperature servo-regulated to a target of 36.8 degrees C. Temperature was recorded throughout the operative period and comparisons were made between groups for both the time and area under the curve (AUC) for a temperature <36 degrees C (AUC<36 degrees C). Twenty-nine patients (AS group = 14, control group = 15) were studied. The AS group had significantly less hypothermia than the control group, both for duration of time <36 degrees C (2.5 [0-22] min, median [interquartile range] AS group versus 118 [49-192] min, control group; P = 0.0008) as well as for AUC<36 degrees C (0.3 [0-2.2] degrees C x min, AS group versus 17.1 [3.6-173.4] degrees C x min, control group; P = 0.002). The Arctic Sun Temperature Management System significantly reduced intraoperative hypothermia during OPCAB surgery. Importantly, this was achieved in the absence of any other temperature modulating techniques, including the use of IV fluid warming or increases in the ambient operating room temperature. The Arctic Sun Temperature Management System was more effective than conventional methods in preventing hypothermia during off-pump coronary artery bypass graft surgery.
Medical subject headings
- Body Temperature
- Coronary Artery Bypass
- Hypothermia
- Intraoperative Complications