Localized hypersensitivity and late coronary thrombosis secondary to a sirolimus-eluting stent: should we be cautious?

Virmani, Renu; Guagliumi, Giulio; Farb, Andrew; Musumeci, Giuseppe; Grieco, Niccolo; Motta, Teresio; Mihalcsik, Laurian; Tespili, Maurizio et al. · Circulation · 2004

case_report · Level V

Where this comes from

Abstract

The US Food and Drug Administration recently issued a warning of subacute thrombosis and hypersensitivity reactions to sirolimus-eluting stents (Cypher). The cause and incidence of these events have not been determined. We present findings of a 58-year-old man who died of late stent thrombosis 18 months after receiving 2 Cypher stents for unstable angina. Although angiographic and intravascular ultrasound results at 8 months demonstrated the absence of neointimal formation, vessel enlargement was present. An autopsy showed aneurysmal dilation of the stented arterial segments with a severe localized hypersensitivity reaction consisting predominantly of T lymphocytes and eosinophils. The known pharmacokinetic elution profile of Cypher stents and the presence of polymer fragments surrounded by giant cells and eosinophils suggest that a reaction to the polymer may have caused late stent thrombosis. Careful long-term follow-up of patients with vessel enlargement after Cypher stent placement is recommended.