Testing of a new pneumatic device to cause pain in humans.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 14766716.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Surgical pain typically combines superficial and deep pain. We wished to generate pain that resembled surgical pain, reliably and reproducibly, in volunteers. We constructed a computer-controlled pneumatic device to apply pressure to the anterior tibia. The reproducibility of the pain was tested by rating the pressure that caused pain rated 4-5 on a visual analogue scale (VAS) on days 0, 7, and 24 in 10 volunteers. The effect of remifentanil (0.025, 0.05, 0.075, and 0.1 microg kg-1 min-1) on pain tolerance in another set of volunteers (n=11) was used as an indirect measure of the reliability of pain production. The pressure needed (0.7 (0.3) to 0.9 (0.4) atm (mean (SD)) to induce pain rated 4-5 (VAS) did not vary, showing long-term reproducibility of the method. When pressure was applied to cause increasing pain in volunteers (n=11) 0.05 microg kg-1 min-1 remifentanil increased pain tolerance by 50%. An approximate doubling of the dose (0.1 microg kg-1 min-1) increased pain tolerance significantly more. The linear logarithmic dose-effect relationship shows that the device causes pain reliably, and this can be reduced with opioid treatment. This pneumatic device can apply pain reliably and reproducibly.
Medical subject headings
- Pain