Left ventricular segmental volume by conductance catheter and Cine-CT.

van der Velde, E T; van Dijk, A D; Steendijk, P; Diethelm, L; Chagas, T; Lipton, M J; Glanz, S A; Baan, J · Eur Heart J · 1992

basic_science · Level V

Where this comes from

Abstract

The ability of the conductance catheter method to measure left ventricular segmental and total volume was evaluated by comparison with the Cine-CT technique. In the seven dogs studied, 19 conductance catheter and simultaneous Cine-CT runs were obtained. High correlation coefficients were found for total volume and segmental volumes, except in the basal segment. However, in most cases there was a significant variability in slope and intercept between animals. Both methods are promising tools for estimating dynamic segmental left ventricular volume, each having specific advantages such as a continuous signal (conductance catheter) or anatomic detail (Cine-CT). However, the results also show the need for further improvement of both methods.