Improvement in intracardiac impedance volumes by field extrapolation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1478207.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The measurement of volume by electrical impedance is complicated by non-homogeneous current distribution resulting from small current sources, by the irregular shape of the ventricle, and by loss of current to surrounding anatomical structures. A mathematical technique, field extrapolation, was developed to partially correct the current distribution. The technique mathematically transforms measured potentials into the potential distribution which would result from infinitely distant current sources. The linear correlation coefficient between impedance stroke volume or cardiac output using field extrapolation and thermodilution stroke volume or cardiac output was 0.83 (n = 86) in 11 dogs and 0.76 (n = 92) in 12 patients. The average linear correlation coefficient between impedance stroke volume and integrated aortic flow in four dogs was 0.83 +/- 0.09 (n = 49) using field extrapolation.
Medical subject headings
- Cardiac Catheterization
- Cardiac Output
- Cardiography, Impedance
- Models, Cardiovascular
- Ventricular Function