Serum ferritin measurement and the degree of agreement using four techniques.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1485604.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Four commercial assay kits to measure serum ferritin were compared to establish the degree of agreement and interchangeability between the different techniques based on the intraclass correlation coefficient (ri). Radioimmunoassay, microparticle enzyme immunoassay, enzyme-linked immunosorbent assay, and chemiluminescent immunoassay systems were used. The Pearson product-moment correlation factor (r) between any two methods was at least 0.98, and the intercept of the regression equations ranged from -0.613 to 3.797, indicating that the methods were comparable. Furthermore, the intraclass correlation coefficient (ri) was at least 0.98, suggesting that the methods were interchangeable.
Medical subject headings
- Enzyme-Linked Immunosorbent Assay
- Ferritins
- Immunoassay
- Immunoenzyme Techniques
- Radioimmunoassay