Clinical evaluation of CAP System and RAST in the measurement of specific IgE.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 1485648.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We investigated the diagnostic value of a new in vitro test, Pharmacia CAP System (Pharmacia Diagnostics AB, Uppsala, Sweden), for the quantitative measurement of allergen-specific IgE antibodies by comparison with RAST in 2 groups of patients, 71 atopic and 48 non-atopic. In the last 20 years RAST has supplied a good diagnostic tool, but this test presents some problems, the main one being sensitivity. The new test has a solid phase able to bind even very small amounts of specific IgE and an anti-IgE tracer with very low cross-reactivity with other immunoglobulins, thus presenting more favourable conditions. From the analysis of our results, Pharmacia CAP System gave higher sensitivity (94% compared to 88% of RAST) with no loss of specificity (96% for both tests). The reliability of these results is ensured by the proper selection of patients who were all suffering from pollinosis and were clinically diagnosed as certainly hypersensitive to a single pollen. A positive trend was found between severity of asthma and levels of specific IgE for timothy. Pharmacia CAP System appears to identify a larger number of atopic patients than RAST.
Medical subject headings
- Antibodies, Monoclonal
- Hypersensitivity, Immediate
- Immunoglobulin E
- Radioallergosorbent Test
- Serologic Tests