Ambulatory oxygen uptake measurement system.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1487290.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An apparatus for monitoring prolonged measurement of oxygen uptake without discomfort has been developed. The system is based on the open flow-through principle. The subject wears a hood through which air is drawn by an exhaust blower. Oxygen concentration is kept constant by a servo-controlled blower, so that flow rate is essentially proportional to oxygen uptake. The flow rate of the main stream is measured by a time-of-flight flowmeter located in the exhaust hose. Oxygen concentration is measured by a limiting-current oxygen sensor. The signals representing flow rate and oxygen concentration are digitized and stored in a memory. All of the equipment is assembled in a small package (wt = 3 kg) which the subject carries on his back. The memory chip is removed from the portable device after each experiment. Data are transferred from the memory to a microcomputer and then an oxygen uptake trend is estimated. Response of the system to a step input is about 30 s. Compared with a commercially available oxygen monitoring system (Magna 88, P. K. Morgan Ltd., Rainham, UK), the maximum difference was 0.15 L/min within an oxygen uptake range of 0.25-2.3 L/min. Thus, the measurements made with our portable device agree with the results obtained using the conventional method.
Medical subject headings
- Monitoring, Physiologic
- Oxygen Consumption