Primary stability of a robodoc implanted anatomical stem versus manual implantation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 14967574.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the initial stability of anatomical stems implanted in manually broached femoral cavities compared with that assessed in cavities milled with the robodoc system. The bone-prosthesis interface motion was measured in matched pairs of cadaveric femora to assess the initial stability of anatomical stems implanted with two different implantation techniques. The high costs of surgical robots and the increased perioperative efforts associated with their use can only be justified if measurable benefits for patients can be achieved. Increased initial stability of the stem as an early indicator for better bone ongrowth would be such a benefit. Seven pairs of fresh frozen human cadaveric femora were used. One femur of each pair was randomly assigned to receive the robotic milling method; the other femur underwent manual broaching by an experienced surgeon. Initial micromotions of the anatomical stems were measured during simulated gait cycles with loads of < or =1500 N, and both groups underwent matched-pair analysis. Results. High motion of the prostheses was found for both implantation techniques. The robodoc system did not enhance the primary stability of the anatomical prosthesis compared with the manual broaching method.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Range of Motion, Articular
- Surgery, Computer-Assisted
Anatomy
- femur
- hip