Validation of coronary flow reserve measurements by thermodilution in clinical practice.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 14972422.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Coronary flow reserve (CFR) and fractional flow reserve (FFR) provide complementary information on the coronary circulation. Using a pressure wire, it is possible to calculate CFR by thermodilution (CFR(thermo)), so that FFR and CFR can be measured with a single guide wire. The present multicentric study was performed to compare the feasibility of CFR(thermo)obtained with an improved algorithm and a standardized injection technique and its agreement with Doppler-derived CFR (CFR(Doppler)). In 86 patients with coronary artery disease recruited during 1 week in eight centres FFR, CFR(thermo)and CFR(Doppler)were measured. FFR could be obtained in all patients (100%). An optimal CFR(Doppler)could be obtained in 69% of the patients. CFR(thermo)could be obtained in 97% of the patients. A significant correlation was found between CFR(Doppler)and CFR(thermo)(r=0.79, P<0.0001) but CFR(thermo)tended to be higher than CFR(Doppler). In a setting close to 'real world' practice, this multicentric study confirms the feasibility and reliability of thermodilution-derived CFR. In addition, the safety and the swiftness of assessing FFR and CFR with one single guide wire makes the latter a unique clinical tool for the evaluation of the coronary circulation.
Medical subject headings
- Coronary Circulation