Mammographic breast glandularity in Malaysian women: data derived from radiography.

Jamal, Noriah; Ng, Kwan-Hoong; McLean, Donald; Looi, Lai-Meng; Moosa, Fatimah · AJR Am J Roentgenol · 2004

cross_sectional · Level IV

Where this comes from

Abstract

This study was undertaken to estimate mammographic breast glandularity in Malaysian women from radiographic data. A mammography X-ray unit was used to expose different thicknesses of phantom material of varying glandular and adipose composition at 27 kV. A least squares method was then used to fit the combined data of phantom glandularity, thickness, and milliampere-seconds. The subsequent fitted equation was then applied to calculate breast glandularity for 705 women who underwent diagnostic mammography, who were drawn equally from the three major ethnic groups of Malaysia: Malay, Chinese, and Indian. The difference in breast glandularity among ethnic groups was tested for significance using the nonparametric Kruskal-Wallis test. The fitted equation gave an absolute error of less than or equal to +/- 8% when applied to the data from phantom exposure. The average breast glandularity of the study sample was 48.9% +/- 18.7%. Breast glandularity was found to decrease with breast thickness and age. No significant difference was seen in breast glandularity among the ethnic groups (p > 0.05, Kruskal-Wallis test).

Medical subject headings