Mammographic breast glandularity in Malaysian women: data derived from radiography.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 14975974.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study was undertaken to estimate mammographic breast glandularity in Malaysian women from radiographic data. A mammography X-ray unit was used to expose different thicknesses of phantom material of varying glandular and adipose composition at 27 kV. A least squares method was then used to fit the combined data of phantom glandularity, thickness, and milliampere-seconds. The subsequent fitted equation was then applied to calculate breast glandularity for 705 women who underwent diagnostic mammography, who were drawn equally from the three major ethnic groups of Malaysia: Malay, Chinese, and Indian. The difference in breast glandularity among ethnic groups was tested for significance using the nonparametric Kruskal-Wallis test. The fitted equation gave an absolute error of less than or equal to +/- 8% when applied to the data from phantom exposure. The average breast glandularity of the study sample was 48.9% +/- 18.7%. Breast glandularity was found to decrease with breast thickness and age. No significant difference was seen in breast glandularity among the ethnic groups (p > 0.05, Kruskal-Wallis test).
Medical subject headings
- Breast
- Mammography