Myocardial viability: breath-hold 3D MR imaging of delayed hyperenhancement with variable sampling in time.

Foo, Thomas K F; Stanley, David W; Castillo, Ernesto; Rochitte, Carlos E; Wang, Yi; Lima, João A C; Bluemke, David A; Wu, Katherine C · Radiology · 2004

case_control · Level III

Where this comes from

Abstract

A method for visualizing myocardial infarction with a three-dimensional (3D) breath-hold gated acquisition was examined. By using variable sampling in time, whole heart coverage with a single volume acquisition was achieved in 24 heart beats. In a study of 35 patients, in whom 3D volume acquisition was compared with a two-dimensional (2D) acquisition, all regions of myocardial infarction were correctly identified at 3D examination. The mean imaging time for 12 section locations was 8.0 minutes +/- 3.0 with a 2D approach compared with 22 seconds +/- 4 with a 3D approach (P <.001). Advantages were also noted for infarct contrast-to-noise ratio: 60 +/- 37 for 3D versus 33 +/- 20 for 2D imaging (P <.001). No significant differences (P >.05) were noted at qualitative assessment of myocardial suppression, endocardial border visualization, respiratory and cardiac motion artifacts, or confidence of transmurality of the infarct.

Medical subject headings