Accurate dose calibrator activity measurement of 90Y-ibritumomab tiuxetan.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 15001686.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This investigation examined the accuracy of dose calibrator activity measurement of the beta-emitting radiopharmaceutical (90)Y-ibritumomab tiuxetan. Five different facilities independently measured (90)Y in a 10-mL syringe geometry with 30 dose calibrator models from 3 different manufacturers. The activities ranged from 81.4 MBq (2.2 mCi) to 1,406 MBq (38 mCi) over the volume range of 3-9 mL. The mean dial settings for (90)Y measurement were 375, 51 x 10, and 897 x 100 for Atomlab, CRC, and Mark V dose calibrators, respectively. The maximum volume dependence was 0.28%/mL. This study demonstrated that when measuring all volumes of (90)Y-ibritumomab tiuxetan activity prescriptions, only a single dial setting for a given manufacturer's dose calibrator is required for accurate measurements. Volume corrections are not necessary. For best accuracy, an individually determined dial value should be used.