Total hip arthroplasty using an Omniflex modular system: 5 to 12 years followup.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 15021139.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated intermediate-term results of primary cementless Omniflex prostheses. Forty-nine patients (57 hips) with a mean age of 44 years were observed for an average of 8.6 years. The average Harris hip score was 46 points before surgery and 87 points at final followup. Revision was done in five hips (9%) in five patients. Two acetabular components (4%) were revised because of breakage of the polyethylene insert. Three femoral components (5%) were revised because of aseptic loosening. Six femoral components (10%) were radiographically loose. Overall, 11 hips (19%) were considered mechanical failures. Osteolysis was identified around three acetabular components (5%) and around 20 femoral components (35%). Nine hips (16%) had femoral osteolysis around the distal stem tip. The average annual polyethylene liner wear was 0.18 mm. These results were inferior to those using other recent cementless total hip systems.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Joint Instability
- Prosthesis Design
Anatomy
- hip
- femur