The Coapsys device to treat functional mitral regurgitation: in vivo long-term canine study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 15052204.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated the capability of the Myocor Coapsys device (Myocor, Inc, Maple Grove, Minn) to reduce functional mitral regurgitation in a canine model of dilated cardiomyopathy. Functional mitral regurgitation with heart failure was induced in 7 dogs by rapid ventricular pacing. The Coapsys device, which consists of anterior and posterior epicardial pads connected by a subvalvular chord, was then implanted. Heart failure was maintained by continued pacing for 8 weeks. Hemodynamic and echocardiographic measurements were performed at pre- and postsizing and after 8 weeks. The Coapsys subvalvular chord was cut to verify that maintenance of valve competency was due to the device. All implants were performed off-pump without atriotomy. Mitral regurgitation was reduced in all animals; mean mitral regurgitation grade was reduced from 2.9 +/- 0.7 to 0.7 +/- 0.8 (P =.00005) and was maintained at 0.8 +/- 0.8 after 8 weeks, without hemodynamic compromise or structural damage to the mitral valve. Mitral regurgitation returned to 3.6 +/- 0.8 (P =.102 versus presizing) after cutting the Coapsys subvalvular chord. The Coapsys device consistently and chronically reduced functional mitral regurgitation. This device is in clinical trials in the United States.
Medical subject headings
- Heart-Assist Devices
- Mitral Valve Insufficiency