Direct computation of ultrasound phased-array driving signals from a specified temperature distribution for hyperthermia.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1505996.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents a new method which obtains ultrasound hyperthermia applicator phased-array element driving signals from a desired temperature distribution. The approach combines a technique which computes array element driving signals from focal point locations and intensities with a new technique which calculates focal point locations and power deposition values from temperature requirements. Temperature specifications appear here as upper and lower bounds within the tumor volume, and a focal point placement algorithm chooses focal patterns capable of achieving the temperature range objective. The linear algebraic structure of the method allows rapid calculation of both the phased-array driving signals and an approximate temperature field response. Computer simulations verify the method with a spherical section array (SSA) for a variety of temperature specifications and blood perfusion values. This scheme, which applies to any phased-array geometry, completes an essential step in both treatment planning and feedback for hyperthermia with ultrasound phased-array applicators.
Medical subject headings
- Computer Simulation
- Hyperthermia, Induced
- Signal Processing, Computer-Assisted
- Thermography
- Ultrasonography