The effect of varying low-dose protocols on perceived image quality in multidetector CT in a rabbit model of acute appendicitis.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 15067426.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to evaluate the value of multidetector CT (MD-CT) with low-dose protocols in the diagnosis of acute appendicitis in a rabbit model. Ten New Zealand white rabbits underwent ligation of the appendix. MD-CT (Siemens Somatom Plus 4 Volume Zoom) was performed using three low-dose protocols: 80 kV/70 mAs; 80 kV/50 mAs; and 80 kV/20 mAs. Axial and coronal reformations were performed. Three experienced radiologists evaluated the images using a four-grade score: (1) excellent; (2) good; (3) moderate; and (4) not sufficient. Diagnosis was confirmed histopathologically. The overall score for the 80 kV/70 mAs was 1, for the 80 kV/50 mAs protocol 1.3, and the 80 kV/20 mAs protocol reached an overall score of 2. The coronal reformations reached with 1.1 (80 kV/70 mAs), 1.2 (80 kV/50 mAs), and 1.4 (80 kV/20 mAs) had a slightly better median score than the axial reconstructions. In cases of inconclusive ultrasound and clinical presentation, the use of MD-CT can be considered when evaluating children for suspected appendicitis.
Medical subject headings
- Appendicitis
- Appendix
- Iohexol
- Tomography, X-Ray Computed