Measurement of density and chemical concentration using a microfluidic chip.

Sparks, D; Smith, R; Straayer, M; Cripe, J; Schneider, R; Chimbayo, A; Anasari, S; Najafi, N · Lab Chip · 2003

basic_science · Level V

Where this comes from

Abstract

A new microfluidic product for measuring fluid density, specific gravity and chemical concentration has been developed. At the core of this lab-on-a-chip sensor is a vacuum-sealed resonating silicon microtube. Measurements can be made with under a microliter of sample fluid, which is over 1000x less than is conventionally required. Since the product is MEMS-based the overall system size is a fraction of conventional density meters and it weighs much less than the traditional desk-top, temperature controlled, density meters. The syringe or pipette loaded system includes a dynamic temperature control system that operates between 0 degree C and 90 degree C with an accuracy of less than 0.01 degree C. Density measurement accuracies of 4 to 5 digits have been observed with aqueous solutions. Measurement examples and applications will be discussed.