Fatal complications after use of the Symmetry Aortic Connector in coronary artery bypass surgery.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 15111196.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
During the last 2 years, 103 aortic saphenous vein graft anastomoses were performed in 68 patients undergoing off-pump coronary artery bypass by using the Symmetry Bypass System Aortic Connector. Of these patients, 2 died during the early postoperative period. In the first patient, after an episode of ventricular fibrillation and closed-chest cardiac massage, the sternum was opened and hemopericardium secondary to leakage of the proximal anastomotic device was found. The second patient died of ascending aortic dissection, the tear of which was likely to have originated from the proximal anastomotic site.
Medical subject headings
- Anastomosis, Surgical
- Coronary Artery Bypass
- Prostheses and Implants