Infected CNS infusion pumps. Is there a chance for treatment without removal?

Boviatsis, E J; Kouyialis, A T; Boutsikakis, I; Korfias, S; Sakas, D E · Acta Neurochir (Wien) · 2004

case_report · Level V

Where this comes from

Abstract

Intrathecal baclofen administration by means of an implantable pump is nowadays a safe and effective method in the treatment of spasticity. One of the rare but devastating complications of this technique is pump infection, with a variety of Gram (-) and Gram (+) organisms being involved. Treatment of these infections, according to international literature, requires removal of the device and appropriate antibiotic therapy. This article reports the authors experience in treating 3 patients with severe, medically intractable spasticity, suffering from infection of the intrathecally-delivering pump. A decision was made not to replace the device, but to treat this complication with pump disinfection and with a new treatment modality that has never been used before, the intra-pocket administration of antibiotics. In all cases the infection was eradicated and the integrity of the pump maintained. None of the patients required a procedure under general anesthesia. Removal should no longer be considered the first treatment option in infections of intrathecally delivering pumps, especially those due to non-adherent bacteria, with mild clinical symptomatology. An initial attempt should always be made for conservative treatment. Intra-pocket administration of antibiotics helps in achieving high drugs levels locally, and may prove an important element in our armamentarium against such infections.