Whole-brain vascular reactivity measured by fMRI using hyperventilation and breath-holding tasks: efficacy of 3D prospective acquisition correction (3D-PACE) for head motion.

Naganawa, Shinji; Koshikawa, Tokiko; Fukatsu, Hiroshi; Ishigaki, Takeo; Maruyama, Katsuya; Takizawa, Osamu · Eur Radiol · 2004

other · Level V

Where this comes from

Abstract

Functional MR imaging (fMRI) study using hyperventilation and breath-holding task has been reported to be one of the non-invasive methods to examine whole-brain vascular reactivity. The purpose of this study was to evaluate the efficacy of a method for 3D prospective detection and correction of head motion (3D-PACE) in a study of whole-brain vascular reactivity using hyperventilation and breath-holding tasks. Eight healthy volunteers were scanned using an fMRI protocol of hyperventilation and breath-holding task blocks at 3 T in separate runs with and without 3D-PACE. In two subjects, two more runs with and without 3D-PACE were repeated. The mean total number of activated voxels +/- standard deviation was 26,405.3+/-1,822.2 in the run with 3D-PACE and 17,329.9+/-2,766.3 in the run without 3D-PACE ( P<0.05), although there is some intersubject variation regarding the effect of 3D-PACE. In the two subjects whose performed two more runs, the number of activated voxels were smaller in the run without 3D-PACE than even in the run with 3D-PACE performed later. We conclude that 3D-PACE is beneficial for fMRI studies of whole-brain vascular reactivity induced by hyperventilation and breath-holding.

Medical subject headings