Accuracy of the Foldes-Biro equation for measurement of oxygen uptake during anaesthesia: a laboratory simulation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 15144292.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Laboratory benchtop testing was conducted of the Biro method for measurement of oxygen uptake from an anaesthetic breathing system by patients during low flow anaesthesia. A high precision flow and gas concentration measurement system was employed in conjunction with a physical gas exchange simulation model to test the theoretical basis of the method under controlled conditions. Simulated oxygen uptake values between 200 and 350 ml x min(-1) were modelled. The Biro-derived measurement of simulated O(2) uptake significantly underestimated the target value (mean difference -88.5 ml x min(-1), or -31.7%). The bias was directly proportional to the simulated O(2) uptake. The Biro method has a systematic bias that cannot be explained by random measurement imprecision.
Medical subject headings
- Anesthesia, Inhalation
- Models, Biological
- Monitoring, Intraoperative
- Oxygen Consumption