Evaluation of capillary collection system for HbA1c specimens.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 1516490.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated the accuracy and stability of a capillary HbA1c collection system for use with a high-performance liquid chromatography analyzer. The collection system requires that 5 ul blood is drawn into a calibrated capillary tube, which is then placed into a vial of stabilizing solution and sent for analysis. The study was conducted on simultaneously drawn capillary and venous blood specimens from 47 pediatric diabetes patients. Accuracy was determined by comparing the capillary to the venous HbA1c values. Stability was measured by analyzing 17 capillary specimens over 3 wk. There was excellent agreement between the capillary and venous HbA1c values (capillary 0.959, venous +0.494, R2 = 98.7%). The capillary HbA1c values were 0.2% higher than the venous HbA1c values and decreased gradually over time (0.1% HbA1c/week) when stored at room temperature. The Bio-Rad (Richmond, CA) collection system is accurate, stable, and simple to use.
Medical subject headings
- Blood Specimen Collection
- Glycated Hemoglobin