Epidural catheter penetration of human dural tissue: in vitro investigation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 15166569.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Factors contributing to subarachnoid catheter passage after epidural placement are not well understood. This study explored mechanisms that might explain its occurrence. Human cadaveric dura was mounted on a model and pressurized to physiologic levels. In a standardized fashion, a 20-gauge Portex three-port, closed end (nonflexible) tip catheter was passed through an epidural needle mounted on a micromanipulator at a 90 degree angle, attempting to penetrate dura with the catheter. Attempts then followed with a 19-gauge Arrow Flex Tip Plus single-port catheter. Subarachnoid catheter passage was compared in (1) intact dura, (2) clinically occult versus obvious epidural needle punctures, and (3) single 25-gauge Whitacre spinal needle punctures after combined spinal-epidural placement. Neither catheter penetrated intact dura: Portex, 0 of 300 attempts (0.0000; 95% confidence interval [CI]: 0.0000, 0.0158); Arrow, 0 of 300 attempts (0.0000; 95% CI: 0.0000, 0.0158). In clinically occult epidural needle punctures, the 20-gauge Portex catheter penetrated 1 of 3 specimens in 1 of 14 attempts (0.0714; 95% CI: 0.0021, 0.3583). The 19-gauge Arrow did not pass (0 of 15 attempts, 0.0000; 95% CI: 0.0000, 0.2535). In clinically obvious epidural needle punctures, the Portex passed in 6 of 33 attempts (0.1818; 95% CI: 0.0760, 0.3608) and the Arrow passed in 1 of 35 attempts (0.0286; 95% CI: 0.0012, 0.1662). Neither catheter passed through a single 25-gauge spinal needle puncture after an uncomplicated combined spinal-epidural: Portex, 0 of 90 attempts (0.0000; 95% CI: 0.0000, 0.0510); Arrow, 0 of 90 attempts (0.0000; 95% CI: 0.0000, 0.0510). Catheter passage is unlikely in the presence of intact dura or after an uncomplicated combined spinal-epidural. Unintentional subarachnoid passage suggests dural damage with the epidural needle.
Medical subject headings
- Catheterization
- Spinal Cord