Accuracy of thin-slice computed tomography in the detection of coronary stenoses.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 15191775.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Our aim was to investigate the accuracy of multislice spiral computed tomography (MSCT) in the detection of significant (>50%) coronary stenosis using a scanner equipped for 16 x 0.625 mm collimation. In 64 patients (59 male, mean age 58 +/- 5 years) with suspected coronary artery disease, MSCT (GE Light Speed-16, collimation: 16 x 0.625 mm) was performed 20 +/- 5 days before coronary angiography (CAG). Only angiographic segments >1.5 mm were considered for analysis. In all patients, MSCT was carried out without complications. Three patients were excluded from the analysis. Of 729 angiographic segments, 613 (84%) were judged evaluable by MSCT. Considering only the segments judged evaluable, the sensitivity was 89%, specificity 98%, positive predictive value 90%, and negative predictive value 98%. Including all segments in the analysis (evaluable and nonevaluable), sensitivity was 78%. Using a scanner with a collimation of 16 x 0.625 mm, our study confirms the potential role of MSCT in the detection of significant coronary stenosis with a sensitivity of 89% and a very high specificity (98%). Exclusion criteria and less than full evaluability of the coronary arteries must still be considered limitations of the method.
Medical subject headings
- Coronary Stenosis
- Tomography, X-Ray Computed