Arterial MR imaging phase-contrast flow measurement: improvements with varying velocity sensitivity during cardiac cycle.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 15220510.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To reduce noise in velocity images of magnetic resonance (MR) phase-contrast measurements, the authors implemented and evaluated a pulse sequence that enables automatic optimization of the velocity-encoding parameter V(enc) for individual heart phases in pulsatile flow on the basis of a rapid prescan. This sequence was prospectively evaluated by comparing velocity-to-noise ratios with those from a standard MR flow scan obtained in the carotid artery in eight volunteers. This sequence was shown to improve velocity-to-noise ratios by a factor of 2.0-6.0 in all but the systolic heart phase and was determined to be an effective technique for reducing noise in phase-contrast velocity measurements.