Delayed biodegradation of a meniscal screw.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 15243418.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A biodegradable polylactide meniscal screw was removed 32 months after the original meniscal repair. It was then subjected to gel permeation chromatography and was found to have retained 64.7% of its original molecular weight. Although it is obvious that the manufacturers of biodegradable implants have succeeded in producing materials that retain their strength long enough to allow tissue healing, the question arises as to the true biodegradation time of the same implants.
Medical subject headings
- Absorbable Implants
- Cartilage Diseases
- Foreign Bodies
- Knee Joint
- Menisci, Tibial
- Tibial Meniscus Injuries
Anatomy
- knee
- tibia