Body conformable 915 MHz microstrip array applicators for large surface area hyperthermia.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 1526638.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The optimal treatment with hyperthermia of superficially located tumors which involve large surface areas requires applicators which can physically conform to body contours, and locally alter their power deposition patterns to adjust for nonuniform temperature caused by tissue inhomogeneities and blood flow variations. A series of 915 MHz microstrip array applicators satisfying these criteria have been developed and clinically tested. Clinical and engineering design tradeoffs for practical devices are discussed. Measurements taken in tissue equivalent phantoms and a summary of our clinical experiences with these microstrip arrays are presented.
Medical subject headings
- Hyperthermia, Induced