Evaluation of Frova, single-use intubation introducer, in a manikin. Comparison with Eschmann multiple-use introducer and Portex single-use introducer.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 15270974.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In a randomised cross-over study, 48 anaesthetists attempted to place a Frova single-use introducer, an Eschmann multiple-use introducer and a Portex single-use introducer in the trachea of a manikin set up to simulate a grade 3 laryngoscopic view. The anaesthetists were blinded to success (tracheal placement) or failure (oesophageal placement). Successful placement (proportion, 95% confidence interval) of either the Frova introducer (65%, 50-77%) or the Eschmann introducer (60%, 46-73%) was significantly more likely than with the Portex introducer (8%, 3-20%). There were no significant differences between the success rates for the Frova and the Eschmann introducers. A separate experiment revealed that the peak force exerted by the Frova and Portex introducers was two to three times greater than that which could be exerted by the Eschmann introducer, p < 0.0001, indicating that the single-use introducers are more likely to cause tissue trauma during placement.
Medical subject headings
- Intubation, Intratracheal
- Manikins