The air elimination capabilities of pressure infusion devices and fluid-warmers.

Schnoor, J; Macko, S; Weber, I; Rossaint, R · Anaesthesia · 2004

Level III

Where this comes from

Abstract

Pressurised infusion devices may have only limited capability to detect and remove air during pressurised infusions. In order to assess pressure infusion systems with regard to their actual air elimination capabilities four disposable pressure infusion systems and fluid warmers were investigated: The Level 1 (L-1), Ranger (RA), Gymar (GY), and the Warmflo (WF). Different volumes of air were injected proximal to the heat exchanger and the remaining amount of air that was delivered at the end of the tubing was measured during pressurised infusions. Elimination of the injected air (100-200 ml) was superior by the RA system when compared to L-1 (p < 0.01). The GY and WF systems failed to eliminate the injected air. In conclusion, air elimination was best performed by the RA system. In terms of the risk of air embolism during pressurised infusions, improvements in air elimination of the investigated devices are still necessary.

Medical subject headings