Three-dimensional lower extremity alignment assessment system: application to evaluation of component position after total knee arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 15284984.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A 3-dimensional (3D) lower extremity alignment assessment system based on biplanar computed radiography, a camera calibration procedure, and 3D image fitting techniques was developed. The goal of this study was to examine the errors associated with applying this 3D technique to determine component alignment after total knee arthroplasty (TKA). Tibial and femoral component alignment after TKA (varus--valgus, flexion-extension, and internal--external angulations) in 4 patients was determined directly with reconstructions from computed tomography (CT) scans and indirectly with the proposed bone and component image fitting techniques. Mean differences between alignment values from the CT and image fitting techniques ranged from 0.18 degrees to 0.78 degrees, and maximum differences ranged from 0.30 degrees to 0.90 degrees. The proposed 3D method can be reliably used to assess component alignment after TKA.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Bone Malalignment
- Imaging, Three-Dimensional
- Knee Joint
Anatomy
- knee