In vivo measurement of swine endocardial convective heat transfer coefficient.

Tangwongsan, Chanchana; Will, James A; Webster, John G; Meredith, Kenneth L; Mahvi, David M · IEEE Trans Biomed Eng · 2004

basic_science · Level V

Where this comes from

Abstract

We measured the endocardial convective heat transfer coefficient h at 22 locations in the cardiac chambers of 15 pigs in vivo. A thin-film Pt catheter tip sensor in a Wheatstone-bridge circuit, similar to a hot wire/film anemometer, measured h. Using fluoroscopy, we could precisely locate the steerable catheter sensor tip and sensor orientation in pigs' cardiac chambers. With flows, h varies from 2500 to 9500 W/m2 x K. With zero flow, h is approximately 2400 W/m2 x K. These values of h can be used for the finite element method modeling of radiofrequency cardiac catheter ablation.