Chemical cytometry on a picoliter-scale integrated microfluidic chip.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 15328405.
- Also identified by PMC identifier 516477.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An integrated microfluidic device has been fabricated for analyzing the chemical contents of a single cell (chemical cytometry). The device is designed to accomplish four different functions: (i) cell handling, (ii) metering and delivering of chemical reagents, (iii) cell lysis and chemical derivatization, and (iv) separating derivatized compounds and detecting them by laser-induced fluorescence. These functions are accomplished with only two valves, formed by multilayer soft lithography. A new kind of three-state valve and a picopipette are described; these elements are crucial for minimizing the reaction volume and ensuring optimal shape of the channel for electrophoresis injection. By using these valves, a reaction volume of approximately 70 pl is achieved for the lysis and derivitization of the contents of a single Jurkat T cell (approximately 10 microm diameter). As a demonstration of the use of this integrated microfluidic device, electropherograms of amino acids from individual Jurkat T cells are recorded and compared with those collected from a multiple-cell homogenate.
Medical subject headings
- Eukaryotic Cells
- Microfluidics
- Nanotechnology