Early tibial tray failure of a Duracon knee with retrieval analysis.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 15343544.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a case of early tibial tray fracture of a Duracon knee prosthesis. Aside from the clinical, radiographic, and gross analysis of the failed prosthesis, we also performed analysis on the retrieved polyethylene component and the fractured tibial baseplate. In the analyses, we noted significant femoral component malalignment, uneven tray cementation, and inherent metallurgical weakness. It appears that the high compressive load on the medial tray resulted in bending fatigue failure. To avoid this complication, it is important to restore the normal alignment of the knee joint and use a polyethylene insert of higher conformity, at least 6 mm thick.
Medical subject headings
- Knee Prosthesis
Anatomy
- knee
- tibia