The effect of tibial implant size on post-operative alignment following medial unicompartmental knee replacement.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 15351414.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We have retrospectively reviewed a consecutive series of 71 medial unicompartmental knee replacements in 56 patients performed between 1991 and 1996 by a single surgeon. The aim of this study was to quantify the amount of correction in the tibiofemoral angle, which is achievable with each thickness of polyethylene tibial insert. Our results have shown that the amount of correction which can be expected with an 8 mm, 10 mm, 12 mm and 14 mm insert are 5.3 degrees, 4.8 degrees, 6.6 degrees and 9.5 degrees, respectively. These results will act as a useful reference for surgeons who perform this procedure.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee
- Knee Prosthesis
- Osteoarthritis, Knee
Anatomy
- knee
- tibia