Performance of the Reflotron in Massachusetts' Model System for Blood Cholesterol Screening Program.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1536371.
- Also identified by PMC identifier 1694384.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The precision, accuracy, and durability of the Reflotron were evaluated by the Massachusetts Model Systems for Blood Cholesterol Screening Program. Screenings were conducted in diverse community settings over 16 months. Fingerstick samples from 10,428 individuals were tested. None of the four analyzers met the 1992 standards for precision, although two met the 1992 standards for accuracy. More than 40% of Reflotron values differed from the reference laboratory values by upwards of 5%. More than 16% of individuals were misclassified in terms of their risk category. All four instruments malfunctioned during the project.
Medical subject headings
- Blood Specimen Collection
- Hypercholesterolemia
- Mass Screening